Analytical Capillary Degasser Chambers

The ”MINI” degassing chamber was designed to be easy to prime, and is configured with a Teflon® AF degassing membrane (tubing) to provide maximum degassing capacity with the absolute minimum internal volume (<3% of PTFE designs with comparable degassing capacity) and a chemically inert, metal free flow path. Degassed mobile phase is available in three minutes or less!
GENERAL SPECIFICATIONS
Degassing Channel Tubing: Teflon® AF
Degassing Channel Pressure Rating: 70 PSIG (priming pressure)
Wetted Materials: Teflon® AF, PEEK and Tefzel
Vacuum Housing Materials: PPS (PolyPhenyleneSulphide) and Stainless Steel

Available Configurations

Systec
Part Number
ApplicationChannel
Volume (μL)
Max HPLC Gradient
Flow CapabilityC (mL/min)
Pressure DropD
(kPa/mL/min)
Degassing Flow
Path ID (mm/in)
9000-1000Capillary1000.40.041.14/0.045
9000-1004Analytical4802.00.181.14/0.045
9000-1006Analytical6702.80.251.14/0.045
9000-1009Analytical9254.00.361.14/0.045
9000-1184Analytical/GPC4802.00.181.14/0.045
* Optimized Flow Range for isocratic or gardient formed 50:50 methanol/water blends. Consult factory for mechanical and chromatographic performance data on various HPLC pumping systems.